Need help? Contact us at (800) 443-8817


Product Listing

Hose Connection

Features
  • Effectively blocks backstreaming by absorbing hydrocarbons
  • Traps water vapor
  • Regeneration using built-in heater
  • Type 304 stainless steel trap body
  • Available in five connection sizes ISO KF and LF ports


Description
Molecular Sieve Traps containing Type 13X synthetic zeolite effectively prevent oil backstreaming at room temperature. They also trap water vapor in the sieve charge. The sieve charge can be reactivated by baking out in place when the base vacuum cannot be achieved. A valve should be used to isolate evolved gas from the system during bakeout cycle. During the bakeout the mechanical pump removes the evolved gas from the sieve trapping material. Normal bakeout can be accomplished by use of a timer during system off-hours.

Any oils trapped by the sieve material cannot be removed by baking. Periodic replacement is required whenever the sieve material exhibits evidence of hydrocarbon loading as determined by empirical observation. Sieve trapping material can be replaced through port on top of trap. Traps can be mounted in the standard vertical or horizontal position.

The built-in heater requires wiring to a 115VAC wall outlet. Once turned on, the heater reaches and maintains a self-regulated constant 150ºC temperature. Bakeout time depends on the amount of water vapor loading of the zeolite, within the range of 4 to 12 hours.

For any question on this Section Ask Our Experts



Drawings download if available are indicated by in the Part Number column. Click on this icon to view all downloads.
Literature downloads (including manuals), if available, is indicated by in the Part Number column. Click on this icon to view all available product literature and to download a PDF copy.

Catalog